Accuracy Specifications |
AC Volts
(40 Hz to 500Hz)1 |
Range |
Resolution |
Accuracy |
6.000 V
60.00 V
600.0 V |
0.001 V
0.01 V
0.1 V |
1.0 % + 3 |
DC Volts |
6.000 V
60.00 V
600.0 V |
0.001 V
0.01 V
0.1 V |
0.5 % + 3 |
AC Millivolts |
600.0 mV |
0.1 mV |
3.0 % + 3 |
Diode Test2 |
2.000 V |
0.001 V |
10 % |
Resistance
(Ohms) |
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ |
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ |
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5% + 3 |
Capacitance3 |
50.00 nF
500.0 nF
5.000 μF
50.00 μF
500.0 μF
1000 μF |
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
1 μF |
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5 |
Frequency4
Hz
(10 Hz – 100 kHz) |
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz |
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz |
0.1 % + 3 |
Duty Cycle4 |
1 % to 99 % |
0.1 % |
1 % typical 5 |
AC Current
(40 Hz to 200 Hz) |
4.000 A
10.00 A |
0.001 A
0.01 A |
1.5 % + 3 |
DC Current |
4.000 A
10.00 A |
0.001 A
0.01 A |
1.5 % + 3 |
- All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
- Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA.
- Specifications do not include
errors due to test lead capacitance and capacitance floor (may be up to
1.5 nF in the 50 nF range).
- All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
- Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %
|
Function |
Overload Protection |
Input Impedance (Nominal) |
Common Mode Rejection Ratio |
Normal Mode Rejection Ratio |
AC Volts |
600 V 1 |
>10 MΩ <100 pF 2 |
>60 dB at dc,
50 Hz or 60 Hz |
− |
AC Millivolts |
600 mV |
>1M, <100 pF |
>80 dB at 50 Hz or 60 Hz |
− |
DC Volts |
600 V 1 |
>10 MΩ <100 pF |
>100 dB at dc,
50 Hz or 60 Hz |
>60 dB at 50 Hz or 60 Hz |
- 6 x 105 V Hz Max
- For mV (AC), input impedance is approximately 1 MΩ.
|
General Specifications |
Maximum Voltage Between any Terminal and Earth Ground |
600 V |
Display (LCD) |
6000 counts, updates 3/sec |
Battery Type |
2 AAA, NEDA 24A, IEC LR03 |
Battery Life |
200 hours minimum |
Temperature |
Operating |
0 °C to 40 °C |
Storage |
-30 °C to 60 °C |
Relative Humidity |
Operating Humidity |
Non-condensing when <10°C
≤90 % at 10 °C to 30 °C; ≤75 % at 30 °C to 40 °C |
Operating Humidity, 40 MΩ Range |
≤80 % at 10 °C to 30 °C; ≤70 % at 30 °C to 40 °C |
Altitude |
Operating |
2000 m |
Storage |
12,000 m |
Temperature Coefficient |
0.1 X (specified accuracy) /°C (<18 °C or >28 °C) |
Fuse Protection for Current Inputs |
11A, 1000V Fast Fuse, Fluke specified part only
|
Size (HxWxL) |
142 mm x 69 mm x 28mm |
Weight |
200 g |
IP Rating |
IEC 60529: IP 40 |
Safety |
IEC 61010-1: 600 V CAT III, Pollution Degree 2 |
Electromagnetic Environment |
IEC 61326-1: Portable |
Electromagnetic Compatibility |
Applies to use the Korea only. Class A Equipment (Industrial Broadcasting and Communication Equipment)1 |